Negative Staining
From emsuite wiki
[edit]
TEM Protocol- Negative Staining
- Need a grid that has formvar and carbon coatings. Ted Pella product number 01801 works very well and is ready to use. 01801 is a copper grid of 200 mesh.
- If the grid package has been opened for a while I recommend that the grids be exposed to U.V. light for no more than 5 minutes.
- Place 10 µl of sample on grid for 2 minutes. Use whatman paper to wick off, be careful to be gentle and only touch the side of the grid.
- Do 2 washes of autoclaved MegPure water. Again use whatman paper to wick off washes.
- Place 10 µ1 1% Uranyl Acetate on grid for 2 minutes. 0.05g of Uranyl Acetate added to 10ml of autoclaved MegPure water.
- Wick off excess stain with whatman paper.
- Air dry.